Background
TactiFlex is a next-generation catheter that is being used increasingly in ablation-treatment strategies. The purpose of this study was to investigate the differences in ablation lesions when the ablation power, time, and perfusion flow are varied with TactiFlex and TactiCath catheters.
Methods
The TactiFlex and TactiCath catheters were contacted perpendicularly/obliquely/parallel to the swine myocardium at varying powers (30, 40, and 50 W), time points (10, 15, 20, 25, 30, and 40 s), and forces (5, 10, 15, 20, and 30 g); the depth, width, and area of each lesion were measured, and the number of steam pops that occurred was counted.
Results
A total of 672 (336 lesions for each catheter) radiofrequency (RF) energy applications were delivered and 648 lesions were analyzed, excluding steam pops. The surface area and volume increased significantly with an increasing contact force for the TactiCath. The TactiCath lesions were significantly deeper than those for TactiFlex in most groups with the same power and seconds. The surface area was significantly larger when the catheters were contacted obliquely/parallel to the myocardium than perpendicularly using TactiCath, whereas the difference was less significant in the case of TactiFlex. In a 10-g contact force ablation experiment, TactiFlex did not cause pops, but TactiCath caused pops in 9.8% of cases.
Conclusion
The TactiFlex lesions were shallower than those of TactiCath. However, TactiFlex catheters reduced the steam pops during RF applications. Those data should help clinicians understand the characteristics of the catheters and develop adequate strategies.