Abstract:Abstracts S129 Results: Figure shows representative waveforms of pump speed, AoP, and pump flow with sinusoidal pump speed modulations of 0%, ±15% and ±25%. The AoP pulse pressure significantly increased to 12.7 mm Hg and 17.7 mmHg with ±15% and ±25% speed modulations, respectively (Table). There were no changes in the mean AoP, mean pump flow, or left atrial pressure. Conclusion: The feasibility to generate pressure pulsatility by pump speed modulation with the CFTAH was validated. This unique platform is sui… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.