OBJECTIVESFrequent screening of SARS-CoV-2 among asymptomatic populations using antigen-based point of care tests (APOCT) is occurring globally with limited clinical performance data. The positive predictive value (PPV) of two APOCT used in the asymptomatic screening of SARS-CoV-2 among healthcare workers (HCW) at continuing care (CC) sites across Alberta, Canada was evaluated.METHODSBetween February 22 and May 2, 2021, CC sites implemented SARS-CoV-2 voluntary screening of their asymptomatic HCW. Onsite testing with Abbott Panbio or BD Veritor occurred on a weekly or twice weekly basis. Positive APOCT were confirmed with a real-time reverse-transcriptase polymerase chain reaction (rRT-PCR) reference method.RESULTSA total of 71,847 APOCT (17,689 Veritor and 54,158 Panbio) were performed among 369 CC sites. Eighty-seven (0.12%) APOCT were positive, of which 39 (0.05%) confirmed as true positives using rRT-PCR. Use of the Veritor and Panbio resulted in a 76.6% and 30.0% false positive detection, respectively (p<0.001). This corresponded to a 23.4% and 70.0% PPV for the Veritor and Panbio, respectively.CONCLUSIONSFrequent screening of SARS-CoV-2 among asymptomatic HCW in CC, using APOCT, resulted in a very low detection rate and a high detection of false positives. Careful assessment between the risks vs benefits of APOCT programs in this population needs to be thoroughly considered before implementation.