2022
DOI: 10.3390/ma15196568
|View full text |Cite
|
Sign up to set email alerts
|

Influence of NiTi Wire Diameter on Cyclic and Torsional Fatigue Resistance of Different Heat-Treated Endodontic Instruments

Abstract: We compared the mechanical properties of 2Shape mini TS2 (Micro-Mega, Besançon, France) obtained from 1.0 diameter nickel-titanium (NiTi) wires and 2Shape TS2 from 1.2 diameter nickel-titanium (NiTi) wires differently thermally treated at room and body temperature. We used 120 NiTi TS2 1.0 and TS2 1.2 files made from controlled memory (CM) wire and T-wire (n = 10). Cyclic fatigue resistance was tested by recording the number of cycles to fracture (NCF) at room and body temperatures using a customized testing d… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

2
6
1

Year Published

2023
2023
2024
2024

Publication Types

Select...
5

Relationship

2
3

Authors

Journals

citations
Cited by 6 publications
(9 citation statements)
references
References 33 publications
2
6
1
Order By: Relevance
“…Comparing instruments with different dimensions, #20.05 files showed higher cyclic fatigue resistance than #25.06, in agreement with previous studies reporting that smaller instruments had high flexibility and fatigue resistance [33,34]. In particular, the HEDM #20.05 reported the highest values of cyclic fatigue resistance.…”
Section: Discussionsupporting
confidence: 88%
“…Comparing instruments with different dimensions, #20.05 files showed higher cyclic fatigue resistance than #25.06, in agreement with previous studies reporting that smaller instruments had high flexibility and fatigue resistance [33,34]. In particular, the HEDM #20.05 reported the highest values of cyclic fatigue resistance.…”
Section: Discussionsupporting
confidence: 88%
“…Therefore, cyclic fatigue tests were conducted in an artificial stainless‐steel canal with the canal dimensions that have been used previously (La Rosa, Palermo, et al, 2021 ; Pedullà et al, 2018 ; Pedullà, Kharouf, et al, 2022 ). Furthermore, the customized apparatus guaranteed standardized laboratory conditions, permitting the reproducibility of the experiments carried out (La Rosa, Palermo, et al, 2021 ; La Rosa, Shumakova, 2021 ; Pedullà, Canova, et al, 2022 ; Pedullà, Kharouf, et al, 2022 ).…”
Section: Discussionmentioning
confidence: 99%
“…The #40/0.03 and #45/0.03 files probably benefit less from this advantage because of their larger size, which negatively impacts their cyclic fatigue behavior. Indeed, the cyclic fatigue is negatively affected by the diameter of the instruments at the point of maximum curvature (Pedullà, Canova, et al, 2022 ). This mechanism occurs because all of its points (except those in the center axis) are subjected to stress/compression cycles during the instrument's rotation.…”
Section: Discussionmentioning
confidence: 99%
See 2 more Smart Citations