1989
DOI: 10.1001/archinte.1989.00390100133028
|View full text |Cite
|
Sign up to set email alerts
|

Initial Clinical Experience With Endocardial Defibrillation Using an Implantable Cardioverter/Defibrillator With a Triple-Electrode System

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

1
7
0

Year Published

1990
1990
2008
2008

Publication Types

Select...
6
2

Relationship

1
7

Authors

Journals

citations
Cited by 79 publications
(8 citation statements)
references
References 16 publications
1
7
0
Order By: Relevance
“…This has been subsequently confirmed during implant procedures for such lead sys tems. Our reported experience and a recent multicenter data analysis indicated highest implant rates (ranging from 69 to 90%) and lowest defibrillation thresholds with the tri ple-electrode configuration [25,26]. In our patients, the mean defibrillation threshold was 18 J.…”
Section: Technologic Advancessupporting
confidence: 47%
See 1 more Smart Citation
“…This has been subsequently confirmed during implant procedures for such lead sys tems. Our reported experience and a recent multicenter data analysis indicated highest implant rates (ranging from 69 to 90%) and lowest defibrillation thresholds with the tri ple-electrode configuration [25,26]. In our patients, the mean defibrillation threshold was 18 J.…”
Section: Technologic Advancessupporting
confidence: 47%
“…The Endotak lead system was as sociated with several incidents of stress frac ture of the conductor coil for both transve nous catheter and patch electrodes [25]. This was manifest as inappropriate shocks, failure to defibrillate at electrophysiologic testing, and even sudden death in one patient.…”
Section: Technologic Advancesmentioning
confidence: 99%
“…1,2 Although present ICD systems using RV leads provide excellent performance, DFTs might be substantially reduced if electric field strength to the LV could be increased. A variety of electrodes, such as subcutaneous patches and arrays, 8,9 active pulse generator "cans," 10 and coils in the CS, 11 have been studied clinically. These approaches typically produce moderate DFT reductions of 20% to 30% and, although many probably increase the electric field strength on the LV, they do not specifically place an electrode in the middle of the LV free wall.…”
Section: Prior Studiesmentioning
confidence: 99%
“…However, their study required the addition of both a CS lead and a subcutaneous patch, did not include the subcutaneous patch (or an active can) as part of the control system, and used 2 biphasic waveforms of equal amplitude (thereby requiring an ICD with 2 sets of discharge capacitors). Because other studies have shown the addition of a subcutaneous patch or active can alone reduces DFTs by 25% to 30%, 8,10 it is unclear how much the addition of a CS lead and sequential shocks might have contributed to the DFT reduction. Our study used a modern, clinically relevant, 3-electrode active can system as a control and showed an additional DFT reduction of 45% using the LV lead and dual-shock waveform.…”
Section: Prior Studiesmentioning
confidence: 99%
“…8 Patients who had this lead suffered multiple inappropriate shocks, and it was withdrawn from clinical trial. Since then, conductor fractures and insulation disruption have challenged engineers and physicians to develop improved designs and implant techniques.…”
Section: Article P 2122mentioning
confidence: 99%