2000
DOI: 10.1016/s1350-4533(00)00081-3
|View full text |Cite
|
Sign up to set email alerts
|

Intramedullary femoral nails: one or two lag screws? A preliminary study

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2

Citation Types

2
29
0
5

Year Published

2011
2011
2024
2024

Publication Types

Select...
7
1

Relationship

0
8

Authors

Journals

citations
Cited by 60 publications
(36 citation statements)
references
References 24 publications
2
29
0
5
Order By: Relevance
“…[20] In our study, the cut-out rate as a result of varus collapse was 4.5% in Group I, consistent with studies by Ruecker et al [21] Zhang et al have reported a cut-out rate of 1-3% in a PFNA group, whereas no patient has developed cut-out in the Intertan group. [19] In the biomechanical study of Huang et al, femoral strength, stability, and bearing capacity have been higher in the InterTAN group than in the PFNA group.…”
Section: Discussionsupporting
confidence: 92%
“…[20] In our study, the cut-out rate as a result of varus collapse was 4.5% in Group I, consistent with studies by Ruecker et al [21] Zhang et al have reported a cut-out rate of 1-3% in a PFNA group, whereas no patient has developed cut-out in the Intertan group. [19] In the biomechanical study of Huang et al, femoral strength, stability, and bearing capacity have been higher in the InterTAN group than in the PFNA group.…”
Section: Discussionsupporting
confidence: 92%
“…The pitfalls and complications were similar and were mainly surgeon or fracture related, rather than implant related. In an experimental study comparing the eYcacy of one versus two femoral neck screws with an intramedullary nail, it was found that a two-lag-screw conWguration should be avoided in patients with a high risk of osteoporotic cut-out [22]. However, lag-screw cut-out remains a signiWcant problem, especially with less stable fracture patterns.…”
mentioning
confidence: 99%
“…Cut-out, Z-effect, reverse Zeffect, lateral migration of the screws are the mostly reported complications resulting from inappropriate placement of screws, low-bone mineral density and alteration of the weight on the lag screws during weight bearing cycles. 9,11,12 Different screw combinations and designs have been developed to overcome these problems and improve the success of treatment.…”
Section: Resultsmentioning
confidence: 99%
“…In a finite element analyses of two lag screws compared with one screw Wang et al had showed that the smaller diameter two screws develop increased stresses in the bone of the femoral head, which potentially could lead to premature screw cut-out. 11 Another issue making treatment of intertrochanteric femur fractures challenging is excessive forces exerted to the femoral head. Cephalomedullary nails have been designed to provide controlled impaction by transforming these forces into a compressor force on fracture site.…”
Section: Resultsmentioning
confidence: 99%