2024
DOI: 10.3390/jcm13082393
|View full text |Cite
|
Sign up to set email alerts
|

Intraoperative Fracture during the Insertion of Advanced Locking Screws (T2 Alpha Femur Retrograde Intramedullary Nailing System): Report of Two Cases and Identifying Causes and Prevention

Takashi Higuchi,
Atsushi Taninaka,
Rikuto Yoshimizu
et al.

Abstract: Background: Recently, the T2 alpha nailing system (Stryker, Inc.), which has advanced locking screws that can attach a screw to a rod, has been used worldwide and is expected to improve fracture fixation. We analyzed two cases of supracondylar femoral fractures in older adult patients, in which intraoperative fractures occurred during the insertion of advanced locking screws of the T2 alpha femur retrograde intramedullary nail. Case presentation: A 93-year-old and an 82-year-old woman each underwent T2 alpha f… Show more