Intraoperative measurement of intraocular pressure using a handheld rebound tonometer with sterilized probes: simple method for intraoperative tonometry

Daniel M. Handzel,
Chiraz Ben Abdallah,
Walter Sekundo
et al.

Abstract: This article describes the utilization of a commercially available device for intraoperative measurement of intraocular pressure (IOP). The presented technique will enable a better understanding of intraoperative dynamics using a commercially available device. 261 phakic eyes were included. Measurements were conducted using the iCare 200. The probes were autoclaved at 134°C for 5 minutes. A successful measurement was achieved in 100% of cases. In 2 cases (0.67%), the initially used probe had to be replaced to … Show more