Abstract: This report covers clinical implementation of a low kV intraoperative radiation therapy (IORT) program with the INTRABEAM® System (Carl Zeiss Meditec AG, Jena, Germany). Based on collective user experience from eight institutions, we discuss best methods of INTRABEAM quality assurance (QA) tests, commissioning measurements, clinical workflow, treatment planning, and potential avenues for research. The guide provides pertinent background information and clinical justification for IORT. It describes the INTRABEA… Show more