Abstract:The work is devoted to the study of atomic vapor absorption spectra obtained in the electrothermal atomizer (graphite furnace) and registered by the high-resolution spectrometer "Grand-2000" (spectral resolution 4 pm). Comparison of the obtained spectra with the spectra recorded by the "Grand-2" spectrometer, which is a part of the "Grand-AAS" atomic absorption spectrometer, indicated the presence of unresolved spectral interferences in the latter, as well as a number of unresolved doublets of analytical lines… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.