Investigation of the dose characteristics of an n-NIPAM gel dosimeter with computed tomography

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
8
0

Year Published

2013
2013
2022
2022

Publication Types

Select...
7

Relationship

2
5

Authors

Journals

citations
Cited by 17 publications
(8 citation statements)
references
References 8 publications
0
8
0
Order By: Relevance
“…The possible reasons are the chemical characteristics of polymer gel dosimeters. According to the previous study [ 3 ], the NIPAM based gel dosimeters have a linear dose response of 2−15 Gy and a minor dose rate dependence of approximately 30%. In the penumbra region of small fields, the dose distribution has a steep slope because of the low scattering compared to one in the central axis.…”
Section: Resultsmentioning
confidence: 99%
See 1 more Smart Citation
“…The possible reasons are the chemical characteristics of polymer gel dosimeters. According to the previous study [ 3 ], the NIPAM based gel dosimeters have a linear dose response of 2−15 Gy and a minor dose rate dependence of approximately 30%. In the penumbra region of small fields, the dose distribution has a steep slope because of the low scattering compared to one in the central axis.…”
Section: Resultsmentioning
confidence: 99%
“…The low-toxicity N -isopropylacrylamide (NIPAM) polymer gel dosimeter proposed by Senden et al [ 1 ] has been intensively researched for the optimization of its formula [ 2 ] and its readout systems. The results showed that the NIPAM gel dosimeter had a high linear dose response ranging from 2–15 Gy [ 3 ], and can be read out by various scanning tools [ 4 , 5 ]. During gel reading, the changes of proton groups in gels can be measured using magnetic resonance imaging (MRI), and the T2 results can be correlated with the 3D dose distribution [ 6 , 7 ].…”
Section: Introductionmentioning
confidence: 99%
“…For the NIPAM-BIS system for example, Huang et al (Huang et al, 2013) reported a decrease of 33% in the dose sensitivity with an increase in the dose rate from 200 cGy/min to 600 cGy/min. On the other hand, Chang et al (Chang et al, 2011a) observed the oposite effect, an increase of almost 33% in the sensititvty when the dose rate was varied from 100 cGy/min to 500 cGy/min, in both cases the experimentes were carried out under 6 MV X-ray energy. Also Senden et al (Senden et al, 2006) …”
Section: Effects Of Dose Ratementioning
confidence: 97%
“…9826, Pyrex, USA) which were subsequently wrapped with tinfoil to prevent ambient lighting and solidified at 4°C for 2 h before use. The n-NIPAM gel has been proven to achieve high linearity and sensitivity under a dose range of 0–20 Gy by CT readouts [31].…”
Section: Methodsmentioning
confidence: 99%
“…One vial of each batch was excluded from irradiation to serve as 0 Gy. After dose delivery, the vials were stored in a 4°C refrigerator for 48 h for polymerization to cease completely [31]. The gel vials were further moved to a −20°C liquid-nitrogen-cooled refrigerator for another 24 h of freeze drying before the gels were cut into 5-mm thin slices alone the long axis for SEM imaging.…”
Section: Methodsmentioning
confidence: 99%