Objectives
Fluorescence resonance energy transfer (FRET)–based ADAMTS13 activity assays are critical for the diagnosis of thrombotic thrombocytopenic purpura. However, these assays are susceptible to interference. As iodide has been suggested to interfere in laboratory testing via fluorophore quenching or promotion, we aimed to determine whether iodinated contrast (Omnipaque) interferes with the ATS-13 ADAMTS13 Activity Assay 2.0.
Methods
We evaluated the excitation, emission, and absorbance spectrum of Omnipaque alone and spiked in patient plasma with known ADAMTS13 activity and ADAMTS13 activity on Omnipaque alone, an abnormal control of patient plasma previously observed to display elevated baseline relative fluorescent units, and variable concentrations of patient plasma with known ADAMTS13 activity spiked with Omnipaque.
Results
No atypical fluorescent peaks were observed on any sample (Omnipaque alone or spiked in plasma) between 250 and 700 nm. There was no difference in the mean ADAMTS13 activity among the various concentrations of plasma spiked with Omnipaque or plasma spiked with saline.
Conclusions
Iodinated contrast does not appear to interfere—either via spectral overlap of the fluorophore or through physiologic inhibition of the ADAMTS13 enzyme—with ADAMTS13 activity FRET-based assays based on the findings from this in vitro analysis. Delaying sample collection for ADAMTS13 activity testing from suspected patients with thrombotic thrombocytopenic purpura following administration of iodinated radiocontrast agents is not necessary, and recent contrast administration should not yield erroneous ADAMTS13 activity results.