238 Pu / Alpha spectrometry / Thermal ionization mass spectrometry / NIST-SRM-947 / Isotopic reference material Summary. A comparative evaluation of 238 Pu determination in NIST-SRM-947-Pu isotopic reference material by alpha spectrometry and thermal ionization mass spectrometry (TIMS) is presented. Electrodeposited sources with geometric progression (G.P.) method to account for tail contribution were used in alpha spectrometry. In TIMS, the recently developed approach of using 235 U as a monitor isotope in interfering element correction methodology was employed to account for the isobaric interference of 238 U at 238 Pu. The atom % abundance of 238 Pu determined by both the methods is in good agreement with each other but is higher than that quoted by NIST in the reference material SRM-947-Pu.