2011
DOI: 10.1088/0031-9155/56/5/012
|View full text |Cite
|
Sign up to set email alerts
|

Iterative deconvolution of simultaneous99mTc and201Tl projection data measured on a CdZnTe-based cardiac SPECT scanner

Abstract: We present a method of correcting self-scatter and crosstalk effects in simultaneous technetium-99m/thallium-201 stress/rest myocardial perfusion SPECT scans. The method, which is in essence a hybrid between the TEW method and scatter modelling, is based on a model of spatial and spectral distribution of projection counts in several selected energy windows. The parameters of the model are determined from measurements of thin rod sources in air when no inobject scatter or attenuation effects are present. The mo… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

1
54
0
2

Year Published

2012
2012
2024
2024

Publication Types

Select...
4
2

Relationship

0
6

Authors

Journals

citations
Cited by 53 publications
(57 citation statements)
references
References 22 publications
1
54
0
2
Order By: Relevance
“…This effect and the pixelated design of dedicated cardiac cameras need to be considered when in-patient scatter and cross-contamination between the photopeaks is to be corrected for. Kacperski et al (2011) has proposed and validated a model-based correction algorithm for dual radionuclide imaging using 99m Tc and 201 Tl for the D-SPECT dedicated cardiac camera with pixelated CdZnTe-detectors. The method employs multiple energy windows and model equations describing the tailing effects and scatter.…”
Section: I-fp-cit (El Fakhri Et Al 2006)mentioning
confidence: 99%
See 4 more Smart Citations
“…This effect and the pixelated design of dedicated cardiac cameras need to be considered when in-patient scatter and cross-contamination between the photopeaks is to be corrected for. Kacperski et al (2011) has proposed and validated a model-based correction algorithm for dual radionuclide imaging using 99m Tc and 201 Tl for the D-SPECT dedicated cardiac camera with pixelated CdZnTe-detectors. The method employs multiple energy windows and model equations describing the tailing effects and scatter.…”
Section: I-fp-cit (El Fakhri Et Al 2006)mentioning
confidence: 99%
“…In this paper we modify the model-based correction algorithm proposed by Kacperski et al (2011) and apply it to the new case of dual radionuclide imaging with 99m Tc and 123 I . Monte Carlo simulations of digital chest phantoms are used in combination with experimental measurement using capillary tubes to estimate the appropriate parameters to be implemented in the algorithm.…”
Section: I-fp-cit (El Fakhri Et Al 2006)mentioning
confidence: 99%
See 3 more Smart Citations