2014
DOI: 10.1088/0026-1394/51/1a/06020
|View full text |Cite
|
Sign up to set email alerts
|

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the ENEA, Italy and the BIPM in medium-energy x-rays

Abstract: A key comparison has been made between the air-kerma standards of the ENEA, Italy and the BIPM in the medium-energy x-ray range. The results show the standards to be in general agreement at the level of the expanded uncertainty (k = 2) of the comparison of 6.2 parts in 10 3. The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2

Citation Types

0
7
0

Year Published

2016
2016
2024
2024

Publication Types

Select...
3

Relationship

1
2

Authors

Journals

citations
Cited by 3 publications
(7 citation statements)
references
References 4 publications
0
7
0
Order By: Relevance
“…All participants had previously participated in the key comparison BIPM.RI(I)-K3 of the airkerma standards of the National Metrology Institutes (NMIs) and the BIPM in mediumenergy X-rays [8][9][10][11]. These results are summarized in table 10.…”
Section: Consistency Check With the Bipmri(i)-k3 Key Comparisonmentioning
confidence: 99%
See 2 more Smart Citations
“…All participants had previously participated in the key comparison BIPM.RI(I)-K3 of the airkerma standards of the National Metrology Institutes (NMIs) and the BIPM in mediumenergy X-rays [8][9][10][11]. These results are summarized in table 10.…”
Section: Consistency Check With the Bipmri(i)-k3 Key Comparisonmentioning
confidence: 99%
“…These results are summarized in table 10. [11]. These values were taken for the consistency check because due to time restrictions ENEA did not use its primary standard but the T4 transfer chamber to calibrate the transfer chambers of this comparison.…”
Section: Consistency Check With the Bipmri(i)-k3 Key Comparisonmentioning
confidence: 99%
See 1 more Smart Citation
“…The x-ray tube is provided with a 3 mm Be window and can be operated at constant potential up to 450 kV in 0.5 kV steps. This x-ray tube is routinely used at ENEA-INMRI for the calibration of instruments with traceability to the Italian national standard of air kerma for medium-energy x-rays (Burns et al 2014). The realization of the radiation qualities used for the work described in this paper is summarized in table 1.…”
Section: Irradiation Facility and Radiation Qualitiesmentioning
confidence: 99%
“…Table 1. Realization of the 180 kV and 250 kV qualities of the medium-energy x-ray CCRI series (Allisy-Roberts et al 2011) at ENEA-INMRI, adapted from (Burns et al 2014). The beam quality is specified also in terms of the copper half value layer (Cu HVL).…”
Section: Irradiation Facility and Radiation Qualitiesmentioning
confidence: 99%