2015
DOI: 10.1016/j.ijgo.2015.04.004
|View full text |Cite
|
Sign up to set email alerts
|

Laboratory evaluation of three dual rapid diagnostic tests for HIV and syphilis in China and Nigeria

Abstract: a b s t r a c t a r t i c l e i n f oKeywords: HIV Rapid tests Sensitivity Specificity SyphilisObjective: To determine the laboratory-based performance and operational characteristics of three dual rapid diagnostic tests (RDTs) for testing HIV and syphilis. Methods: Three dual RDTs (SD Bioline, Chembio, and MedMira) were evaluated using 1514 serum specimens archived at laboratories or collected from clinics in China and Nigeria to determine sensitivity and specificity, with 95% confidence intervals. Concordanc… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1

Citation Types

2
30
0

Year Published

2015
2015
2020
2020

Publication Types

Select...
7
1

Relationship

0
8

Authors

Journals

citations
Cited by 26 publications
(32 citation statements)
references
References 15 publications
2
30
0
Order By: Relevance
“…For Treponema pallidum antibody detection, the electronic reader with the default cutoff value of 10 had the best sensitivity and specificity. Our results are comparable to those of laboratory evaluations that were conducted previously with different point-of-care tests for HIV and treponemal antibody detection in sera and whole blood, which also yielded high performance values for sensitivity and specificity (1,(5)(6)(7)(11)(12)(13). In our evaluation, the sensitivity of the Chembio DPP HIV-syphilis rapid test to detect HIV antibodies was somewhat greater than its sensitivity to detect syphilis antibodies, in concurrence with several other studies (5,6,11,12,14).…”
supporting
confidence: 90%
See 1 more Smart Citation
“…For Treponema pallidum antibody detection, the electronic reader with the default cutoff value of 10 had the best sensitivity and specificity. Our results are comparable to those of laboratory evaluations that were conducted previously with different point-of-care tests for HIV and treponemal antibody detection in sera and whole blood, which also yielded high performance values for sensitivity and specificity (1,(5)(6)(7)(11)(12)(13). In our evaluation, the sensitivity of the Chembio DPP HIV-syphilis rapid test to detect HIV antibodies was somewhat greater than its sensitivity to detect syphilis antibodies, in concurrence with several other studies (5,6,11,12,14).…”
supporting
confidence: 90%
“…Dual screening tests for antibodies to HIV and Treponema pallidum have been evaluated in serum and venipuncture whole-blood samples, showing excellent sensitivity and specificity (5)(6)(7). We assessed the laboratory performance of the Chembio dual-path platform (DPP) HIV-syphilis rapid test and the Chembio digital electronic reader with stored serum samples collected from a cohort of men who have sex with men (MSM) and transwomen at high risk for HIV and syphilis infections.…”
mentioning
confidence: 99%
“…Previous studies evaluated dual tests for antibodies to HIV and T. pallidum, including the MedMira Multiplo (MedMira; Halifax, Nova Scotia, Canada), the SD BIOLINE HIV/Syphilis Duo test (Standard Diagnostics; Gyeonggi-do, Republic of Korea), and the Chembio DPP HIV-syphilis assay (Chembio Diagnostics; Medford, NY) (7,(14)(15)(16)(17)(18)(19)(20)(21)(22)(23). These dual tests demonstrated sensitivities and specificities for HIV antibody detection ranging from 93.8% to 100% and 91.9% to 100%, respectively.…”
mentioning
confidence: 99%
“…In those studies, sensitivities of the POCTT varied from around 63 % in the first studies to almost 100 % [30][31][32][33]36,39,44,45], with specificities varying from around 90 % to 100 % in the most recent studies. The tests' increasing performance over the years is associated not only with the development of better tests but also with better quality control, one of the most important steps, especially when taking into consideration that these tests are to be used mainly in remote regions.…”
Section: Poc Tests' Performancementioning
confidence: 98%
“…Dual [25,36,39], triple [40][41][42], or quadruple [43] POCT for syphilis, HIV, and other STIs have recently been developed, and, for syphilis POCT alone, depending on the conditions and populations studied, quality control implementation and integration in local health facilities have shown to have good sensitivity and specificity for the tests included in the device.…”
Section: Syphilis Serological Diagnosis Algorithmsmentioning
confidence: 99%