Laboratory Testing of Leica At401 Laser Tracker

Abstract: Abstract. This paper describes laboratory tests on a Leica AT401laser tracker. As the newer Leica AT402 model also uses the same firmware package, most of the results should also be valid for this device. First, we present the instrument's firmware errors and the software used for testing. The ASME B89.4.19-2006 standard for testing laser trackers is briefly presented. The warm-up effect of the instrument is inspected with respect to both angle measurement and distance measurement. The absolute distance meter … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

1
9
0

Year Published

2016
2016
2021
2021

Publication Types

Select...
5
1
1

Relationship

2
5

Authors

Journals

citations
Cited by 9 publications
(10 citation statements)
references
References 6 publications
1
9
0
Order By: Relevance
“…Experimental studies have shown that the values of the absolute error in measuring the length and volumetric measurements, determined using the model of the stand, do not exceed the permissible values established when the type of the absolute tracker Leica AT401 was approved [7]. Thus, a model of a stand for metrological support of absolute trackers has been proposed and developed, a method for assessing the metrological characteristics of absolute trackers has been proposed and tested.…”
Section: Discussionmentioning
confidence: 99%
“…Experimental studies have shown that the values of the absolute error in measuring the length and volumetric measurements, determined using the model of the stand, do not exceed the permissible values established when the type of the absolute tracker Leica AT401 was approved [7]. Thus, a model of a stand for metrological support of absolute trackers has been proposed and developed, a method for assessing the metrological characteristics of absolute trackers has been proposed and tested.…”
Section: Discussionmentioning
confidence: 99%
“…Thermal expansion caused by the process of internal heating has a significant impact on the measurement. There is also information from the Leica geosystem given in [4] that the inclination sensor in the oil bath suffers from heating as well and changes its zero position. According to the manufacturer, this is the major contributor to the warm-up effect.…”
Section: Warm-up Effectmentioning
confidence: 99%
“…It specifies a period after which LT reaches a stable internal temperature level. This value also depends on the model of LT, but research has shown that it takes around two hours to reach a stable temperature for different LT [4], [5], [6]. Manufacturers are providing more optimistic information.…”
Section: Warm-up Effectmentioning
confidence: 99%
“…The author of this paper has tested the Leica AT401 in laboratories of Research Institute of Geodesy, Topography and Cartography (RIGTC) and Czech Technical University in Prague, Faculty of Civil Engineering. The main purpose was to prepare the instrument and other equipment for high accurate field measurements at Czech national field length baselines [21].…”
Section: Refraction Model Of the Leica At401mentioning