Leadless or transcatheter pacemakers have recently been introduced to market with important benefits and some limitations. Implanted entirely within the right ventricle, these devices eliminate the need for transvenous pacing leads and pacemaker pockets and thus reduce the risk of infections and lead-related problems. Currently, they offer only VVI/R pacing and they cannot provide atrial sensing, antitachycardia pacing, or AV synchrony. They offer a number of features (such as rate response) and electrogram storage, albeit more limited than in a transvenous system. Real-world clinical data are needed to better comment on projected battery life, which manufacturers suggest will be at least equivalent to transvenous devices. Extracting an implanted leadless pacemaker remains a challenge, although proprietary snare and removal systems are available. However, a leadless pacemaker at end of service may be programmed to OOO and left in place; a revised device may be implanted adjacent. These innovative new devices may have important uses in special populations. Initial data on implant success and adverse events are favorable. Currently, there are two leadless pacemakers available: the Micra™ device by Medtronic and the Nanostim™ device by Abbott (formerly St. Jude Medical).