Abstract:A recently published case documented left ventricular (LV) inhibition of a Boston Scientific device by a premature complex (VPC) that was undetected by the right ventricular channel. We have observed a similar response in two patients with a BIOTRONIK cardiac resynchronization device also designed with LV sensing. A spurious response simulating that of the two true cases was also observed in a third patient with a defective LV lead which created isolated false signals. The responses of the BIOTRONIK devices we… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.