2015
DOI: 10.1016/j.cca.2015.03.044
|View full text |Cite
|
Sign up to set email alerts
|

Liver function testing on the Abaxis Piccolo Xpress: Use in Ebola virus disease protocols

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

3
11
0

Year Published

2015
2015
2022
2022

Publication Types

Select...
5
1

Relationship

0
6

Authors

Journals

citations
Cited by 15 publications
(14 citation statements)
references
References 10 publications
3
11
0
Order By: Relevance
“…Few instruments available in the market provide liver function panels or lipid panel analysis; therefore, studies evaluating these devices are scarce. As expected, the detected levels of total protein were higher in LiHep whole blood than serum; this was attributed to the consumption of coagulation factors and fibrinogen during clotting . Correlation coefficients for liver function tests were similar to those reported by previous studies that compared measurements of LiHep whole blood, performed using the Abbott Piccolo Xpress, to those obtained in a central laboratory .…”
Section: Discussionsupporting
confidence: 87%
See 2 more Smart Citations
“…Few instruments available in the market provide liver function panels or lipid panel analysis; therefore, studies evaluating these devices are scarce. As expected, the detected levels of total protein were higher in LiHep whole blood than serum; this was attributed to the consumption of coagulation factors and fibrinogen during clotting . Correlation coefficients for liver function tests were similar to those reported by previous studies that compared measurements of LiHep whole blood, performed using the Abbott Piccolo Xpress, to those obtained in a central laboratory .…”
Section: Discussionsupporting
confidence: 87%
“…As expected, the detected levels of total protein were higher in LiHep whole blood than serum; this was attributed to the consumption of coagulation factors and fibrinogen during clotting . Correlation coefficients for liver function tests were similar to those reported by previous studies that compared measurements of LiHep whole blood, performed using the Abbott Piccolo Xpress, to those obtained in a central laboratory . However, when the % bias was compared, the data for amylase and GGT showed lower bias while those for the remaining analytes (AST, ALT, ALP, albumin, and total protein) showed higher bias than in other studies .…”
Section: Discussionsupporting
confidence: 85%
See 1 more Smart Citation
“…Using centrifugal and capillary forces, plasma separation, metering, dilution, mixing, and distribution to the cuvettes containing the lyophilised reagents take place in the disc. The Piccolo has been used by the French, British, and US Military in remote locations and has also proved to be an appropriate blood chemistry analyser for biosafety environments, such as those required for monitoring Ebola patients [ 44 ]. Claros Diagnostics (now part of OPKO Health) is another company that has focused its efforts on the development of POC diagnostic devices for developing countries.…”
Section: The Future: Microfluidics As An Emerging Platform For Poimentioning
confidence: 99%
“…accuracy, precision, specimen stability, consistency etc.) of the Piccolo Xpress analyzer have been investigated in adult and neonatal intensive care unit populations [4] , [5] , [6] , [7] . Our study adds to the literature by providing additional comprehensive evaluation in the pediatric population and interference studies using the Liver Panel Plus discs.…”
Section: Introductionmentioning
confidence: 99%