Dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) is widely used to diagnose cancer and monitor therapy. The maximum enhancement ratio (ER max ) obtained from the curve of signal intensity over time could be a biomarker to distinguish cancer from normal tissue or benign tumors. We evaluated the impact of flip angle (FA) and repetition time (TR) on the ER max values of dynamic gadobutrol-enhanced MR imaging, obtaining T 1 -weighted (T 1 W) MR imaging of VX2 tumors using 2-dimensional fast spoiled gradient echo (2D FSPGR) with various FAs (30°, 60°and 90°) at 1.5 tesla before and after injection of 0.1 mmol/kg gadobutrol. In vivo study indicated significant differences between ER max values and area under the ER-time curve (AUC 100 ) of VX2 tumors and muscle tissue, with the highest ER max and AUC 100 at FA 90°. Computer simulation also demonstrated the ER as a strictly increasing monotonic function in the closed interval [0°, 90°] for a given TR when using T 1 W FSPGR, and the highest ER value always occurred at FA 90°. The FA for the highest ER differed from that for the highest signal-to-noise or contrast-to-noise ratio. For long TR, the ER value increases gradually. However, for short TR, the ER value increases rapidly and plateaus so that the ER value changes little beyond a certain FA value. Therefore, we suggest use of a higher FA, near 90°, to obtain a higher ER max for long TR in 2D SPGR or FSPGR and a smaller FA, much less than 90°, to reach an appropriate ER max for short TR in 3D SPGR or FSPGR. This information could be helpful in setting the optimal parameters for DCE-MRI.