“…Rodent hearts were also imaged using FCM and a method of local dye delivery based on a dye carrier affixed to the tip of the FCM imaging microprobe. Dye carriers were fabricated according to a recently described method [ 15 ] and loaded for 15 min with fluorescent dye, sodium fluorescein [Fluorescite® (fluorescein injection, USP) 10%; Alcon, Fort Worth, TX, US; 1:1000]. Images from AWM, SAN, and AVN regions of perfused rodent hearts were acquired using this method with a custom fiber-optics microprobe having a numeric aperture of 0.8 (UltraMiniOWD30; Mauna Kea Technologies) at a lateral resolution ( xy ) of 1.4 μm, optical sectioning ( z ) of 7 μm, xy field of view of 186 by 130 μm, frame rate of 12 Hz, and a z-scan range of 26 μm.…”