Purpose
To test a stent-graft specifically designed for the ascending aorta in phantom, cadaver, and clinical application, and to measure deployment accuracy to overcome limitations of existing devices.
Methods
A stent-graft has been designed with support wires to fixate the apices toward the inner curvature, thereby eliminating the forward movement of the proximal end which can happen with circumferential tip capture systems. The device was deployed in three aortic phantoms, and in four cadavers, deployment precision was measured. Subsequently, the device was implanted in a patient to exclude a pseudoaneurysm originating from the distal anastomosis after ascending aortic replacement.
Results
The stent-grafts were successfully deployed in all phantoms and cadavers. Deployment accuracy of the proximal end of the stent-graft was within 1 mm proximally and 14 mm distally to the intended landing zone on the inner curvature, and 2–8 mm distal to the intended landing zone on the outer curvature. In clinical application, the pseudoaneurysm could be successfully excluded without complications.
Conclusion
The novel stent-graft design promises accurate placement in the ascending aorta. The differential deployment of the apices at the inner and outer curvatures allows deployment perpendicular to the aortic axis.
Level of Evidence
No level of evidence.