Abstract:Assessing accuracy of radiation counting systems over time is critical. We examined long-term WBC performance in detail. Efficiency factors for 54 detectors were updated annually over several years. Newer efficiency values were compared with baseline and with annual values. Overall system efficiency has declined (-1.9% over 3 yrs) and appears to be doing so at an increasing rate. Having more specific performance data on individual components can make the process of system maintenance and repair more straightfo… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.