“…More recently, a new cautery-enhanced LAMS (Hot-Spaxus; Taewoong Medical) has been designed with rounded edges and foldable flanges to potentially minimize bleeding risk. To ascertain the benefit of this, an international team of collaborators conducted a retrospective propensity study to compare the technical and clinical performance, in particular the bleeding risk, of the Hot-Axios and Spaxus (cold and hot) systems [5]. Among the 264 analyzed patients (132/cohort), while there was no difference in the technical or clinical outcomes, the rates of overall adverse events (AEs; 9.8% vs. 3.0%; P = 0.04) and clinically relevant bleeding (6.8% vs. 1.5%; P = 0.03) were significantly higher in patients treated using the Hot-Axios system.…”