2013
DOI: 10.1016/j.ultramic.2012.11.003
|View full text |Cite
|
Sign up to set email alerts
|

Measurement of geometrically necessary dislocation density with high resolution electron backscatter diffraction: Effects of detector binning and step size

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

4
139
0

Year Published

2015
2015
2023
2023

Publication Types

Select...
5
1
1

Relationship

0
7

Authors

Journals

citations
Cited by 234 publications
(143 citation statements)
references
References 33 publications
4
139
0
Order By: Relevance
“…The lowest GND density readily detected by conventional EBSD has been assessed by various authors (Adams and Kacher, 2010;Chekhonin et al, 2014;Field et al, 2012) and is step size dependent. In copper at a 0.5µm step size GND density below 10 14 m -2 cannot be determined using conventional EBSD (Adams and Kacher, 2010;Jiang et al, 2013b). Mishin et al (Mishin et al, 2009) suggested the emergence of HR-EBSD technique might provide promising deformation structure results for a much larger range of plastic deformation.…”
Section: Development Of High Resolution Electron Backscatter Diffractmentioning
confidence: 99%
See 3 more Smart Citations
“…The lowest GND density readily detected by conventional EBSD has been assessed by various authors (Adams and Kacher, 2010;Chekhonin et al, 2014;Field et al, 2012) and is step size dependent. In copper at a 0.5µm step size GND density below 10 14 m -2 cannot be determined using conventional EBSD (Adams and Kacher, 2010;Jiang et al, 2013b). Mishin et al (Mishin et al, 2009) suggested the emergence of HR-EBSD technique might provide promising deformation structure results for a much larger range of plastic deformation.…”
Section: Development Of High Resolution Electron Backscatter Diffractmentioning
confidence: 99%
“…This problem was overcome by designing a Matlab software script (http://www.mathworks.co.uk/) to transfer and store all EBSD images into a single HDF5 file (http://www.hdfgroup.org/) (Jackson et al, 2014;Jiang et al, 2013b). This allowed the ~16 frames per second rate to be maintained throughout the data acquisition process.…”
Section: Mechanical Test and Ebsdmentioning
confidence: 99%
See 2 more Smart Citations
“…Subsequent analysis to calculate residual stress distributions and stored GND density was performed offline using the HR-EBSD analysis method described elsewhere [40][41][42][43][44]. Briefly, HR-EBSD involves direct pattern comparison of two or more diffraction patterns to determine the lattice rotation and residual elastic strain gradients between the crystals which are illuminated by the electron beam.…”
Section: Ebsd and Hr-ebsd Measurementmentioning
confidence: 99%