Measurement of NO<sub><i>x</i></sub> and NO<sub><i>y</i></sub> with a thermal dissociation cavity ring-down spectrometer (TD-CRDS): instrument characterisation and first deployment

Abstract: Abstract. We present a newly constructed, two-channel thermal dissociation cavity ring-down spectrometer (TD-CRDS) for the measurement of NOx (NO+NO2), NOy (NOx+HNO3+RO2NO2+2N2O5 etc.), NOz (NOy−NOx) and particulate nitrate (pNit). NOy-containing trace gases are detected as NO2 by the CRDS at 405 nm following sampling through inlets at ambient temperature (NOx) or at 850 ∘C (NOy). In both cases, O3 was added to the air sample directly upstream of the cavities to convert NO (either ambient or formed in the 850 … Show more

Help me understand this report
View preprint versions

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
24
0

Year Published

2021
2021
2023
2023

Publication Types

Select...
4
1
1

Relationship

2
4

Authors

Journals

citations
Cited by 15 publications
(24 citation statements)
references
References 99 publications
0
24
0
Order By: Relevance
“…frequent occurrence of NOx plumes resulting from emissions of nearby ships. The NOx mixing ratios measured by TD-CRDS and CLD were in good agreement (see Friedrich et al (2020)) and the bias of the TD-CRDS to higher values reflects the exclusion of data below the detection limit. A histogram of the NOx measurements made by the CLD is displayed in Fig.…”
Section: Noxmentioning
confidence: 61%
“…frequent occurrence of NOx plumes resulting from emissions of nearby ships. The NOx mixing ratios measured by TD-CRDS and CLD were in good agreement (see Friedrich et al (2020)) and the bias of the TD-CRDS to higher values reflects the exclusion of data below the detection limit. A histogram of the NOx measurements made by the CLD is displayed in Fig.…”
Section: Noxmentioning
confidence: 61%
“…The TD-CRDS instrument, its operating principles and laboratory characterisation, and a validation of the NOx measurements versus an independent CLD instrument have recently been presented (Friedrich et al, 2020). The TD-CRDS (located in an airconditioned research container on the front deck of the vessel) has two separate cavities operating at a wavelength of 405 nm and at sub-ambient pressure (720 to 770 hPa) to prevent condensation of water in inlet lines under humid conditions.…”
Section: Td-crds Instrument For Nox Noy and Noz Detectionmentioning
confidence: 99%
“…One of the heated inlets was equipped with an activated carbon denuder in order to remove gas-phase NOy species from the sample stream and thus, in principal, was used to measure only particulate nitrate. The oven connected to the inlet featuring the denuder malfunctioned very early in the campaign so that TD-CRDS measurements of particulate nitrate are not available (a short time frame is discussed in Friedrich et al (2020)). In the Red Sea and Arabian Gulf, the oven heating the inlet of the total-NOy channel was also switched off occasionally during the hottest hours of the day, to prevent damage to the oven electronics.…”
Section: Td-crds Instrument For Nox Noy and Noz Detectionmentioning
confidence: 99%
See 2 more Smart Citations