The problem of ensuring the accuracy and traceability of the measurement results of the absorbed dose in carbon ion beams, as well as the measurement results of the amount, fluence, flux density and energy of particles in proton and heavy charged particles beams is considered. Until now, in practice, these values have been measured only by indirect methods. The lack of approved measuring instruments for the quantities under consideration and the metrological traceability of measurement results of these quantities to standards did not allow achieving consistency of measurement methods used in practice and confirming the reliability of the results obtained. To solve this problem, three measuring complexes have been developed and created, which are included in the State Primary Standard of units of absorbed dose and absorbed dose rate of photon, electron, proton radiation and in carbon ion beams, quantity, fluence, flux density and energy of particles in proton and heavy charged particles beams GET 38-2024. The measuring complex for reproducing the unit of absorbed dose in carbon ion beams consists of an adiabatic calorimeter, a thermostating system, a data collection and processing system and a vacuum pumping station. To reproduce the unit of energy of protons and heavy charged particles, a complex has been implemented, which includes a total absorption calorimeter, a data collection and processing system, a vacuum pumping station and a particle count determination system based on the use of a Faraday cup. To reproduce the units of fluence and particle flux density in proton and heavy charged particles beams, a measuring complex has been created containing a Faraday cup, a set of collimators and a low current meter. The schemes, the principles and the results of studies of the metrological characteristics of the developed measuring complexes are described. The results are relevant for the field of radiation therapy and radiation resistance tests of the electronic component base used in the space industry.