Background:
Due to advances in technology and medical devices, intra-thoracic left ventricular assisted devices such as the fully magnetically levitated centrifugal-flow pump may now prolong the life of patients with advanced heart failure. However, several concerns have been raised about pump thrombosis and durability of the device. We aimed to systematically compare the two year outcomes of magnetic levitated centrifugal continuous flow circulatory pump versus the axial continuous flow pump for advanced heart failure.
Methods:
Following the PRISMA guideline, online databases were searched for relevant trials based on centrifugal continuous flow circulatory pump and axial continuous flow pump in patients with advanced heart failure. The adverse clinical outcomes reported at 2 years follow-up were considered as the endpoints. This analysis was carried out by the RevMan 5.3 software whereby odds ratios (OR) and 95% confidence intervals (CI) were generated.
Results:
A total number of 1011 patients with advanced heart failure was included. At 2 years, pump thrombosis was not significantly different between the two groups, with OR: 0.43, 95% CI: 0.06–3.29;
P
= .42. However, pump replacement was significantly higher with the axial continuous-flow pump with OR: 0.36, 95% CI: 0.15–0.84;
P
= .02. Stroke, sepsis and bleeding events were not significantly different. In addition, outcomes such as right heart failure, cardiac arrhythmia, the need for right ventricular assisted device, respiratory failure, renal failure and hepatic dysfunction were also not significantly different.
Conclusions:
At a follow-up time period of 2 years, pump replacement was significantly higher with the axial continuous-flow pump in comparison to the magnetic levitated centrifugal continuous flow circulatory pump. However, no significant difference was observed with the other adverse outcomes.