A mass airflow cabinet for handling and housing of laboratory rodents has been developed and tested. The unit consists of a high-efficiency particulate air filter and uniform distribution of air at a vertical velocity of 19 cm per s. Animals are maintained without bedding in mesh-bottomed cages that rest on rollers for rotation inside the cabinet. There is an air barrier of 90 cm per s separating the cabinet air from room air. Sampling for airborne bacteria yielded an average of 0.03 colony-forming units (CFU) per ft3 of air inside the cabinet, whereas 28.8 CFU per ft8 was simultaneously detected outside the cabinet during housekeeping, a reduction of almost three logs. The efficiency of the air barrier was tested by aerosolization of T3 phage. When phage was aerosolized 5 cm outside the cabinet, no phage could be detected 5 cm inside when the fans were operating; with the fans off an average of 1.6 x 101 plaque-forming units (PFU) per ft3 was