Introduction: EBUS is a well established minimally invasive diagnostic tool for mediastinal and hilar lymphadenopathy. The novel ViziShot Flex 19G needle (Olympus Respiratory America, Redmond, WA, USA) was introduced in 2015 in order to improve loaded scope flexion and to obtain larger tissue samples for analysis. The aims of this study were to assess diagnostic yield of Flex 19G needles and to present endoscopist's feedback about the endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA).
Material and methods:The Flex 19G needles were used in patients with hilar and/or mediastinal adenopathy in two Polish pulmonology centers. Cytology smears and cell blocks (CB) were prepared. The prospective analysis was performed due to collected data. Results: Twenty two selected patients with confirmed adenopathy on chest-CT (mean age 58 ± 12) underwent EBUS-TBNA with use of Flex 19G needles. All procedures occurred to be diagnostic for smears (yield 100%). The malignancy was found in 15 cases (68.2%), and benign adenopathy in 7 (31.8%). In 12 of 14 cases of lung cancer (yield 85.7%) CB were diagnostic for immunohistochemical and molecular staining. After puncturing nodes, especially in hilar position not extensive bleeding was observed. Comparing to standard 21/22G EBUS-TBNA endoscopists underlined better flexion of loaded scope and sample adequacy and found non-significant differences in another biopsy details.
Conclusions:The first Polish experience with use of Flex 19G EBUS-TBNA needle occurs to be similar in performance with standard technique with use of 22/21G needles and presents high diagnostic yield for lung cancer diagnostics, especially when preparing CB. A safety profile of the biopsy is acceptable.