“…The iron-rhodium thermometer used was calibrated by IST-90. The reliability of the calorimeter operation was tested by measuring the heat capacities of specialpurity copper (OSCh 11-4), standard synthetic corundum [5] and K-1 benzoic acid [5] prepared at metrological institutions of the State Standard Committee of Russian Federation. It was established that the apparatus and the measurement procedure allowed us to obtain the C p;m values of the substances in a condensed state with an uncertainty of not more than 2% from T = (6 to 10) K, 0.5% between T = (10 and 40) K, and within 0.2% in the range T = (40 to 355) K.…”