Microbiological cleaning and disinfection efficacy of a three-stage ultrasonic processing protocol for CAD-CAM implant abutments

Abstract: PURPOSE Computer-aided design and manufacturing (CAD-CAM) of implant abutments has been shown to result in surface contamination from site-specific milling and fabrication processes. If not removed, these contaminants can have a potentially adverse effect and may trigger inflammatory responses of the peri-implant tissues. The aim of the present study was to evaluate the bacterial disinfection and cleaning efficacy of ultrasonic reprocessing in approved disinfectants to reduce the microbial load of… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3

Citation Types

2
1
0

Year Published

2023
2023
2024
2024

Publication Types

Select...
5
1

Relationship

0
6

Authors

Journals

citations
Cited by 6 publications
(3 citation statements)
references
References 41 publications
2
1
0
Order By: Relevance
“…The most significant difference in the contamination level between the FP/NC, CP/NC, UD/NC and PL/NC groups should be highlighted. The results are consistent with Canullo et al and Gehrke et al, who also reported superior ultrasonic cleaning for residual and bacterial decontamination [ 2 , 5 , 37 ]. The most effective surface cleanliness for ultrasonic cleaning was measured by Canullo et al at 0.0007%; generally higher contamination levels were measured for two-piece abutments as detected by Gehrke et al, even with multi-stage ultrasonic cleaning, which is also consistent with the current measurements [ 2 , 5 ].…”
Section: Discussionsupporting
confidence: 91%
“…The most significant difference in the contamination level between the FP/NC, CP/NC, UD/NC and PL/NC groups should be highlighted. The results are consistent with Canullo et al and Gehrke et al, who also reported superior ultrasonic cleaning for residual and bacterial decontamination [ 2 , 5 , 37 ]. The most effective surface cleanliness for ultrasonic cleaning was measured by Canullo et al at 0.0007%; generally higher contamination levels were measured for two-piece abutments as detected by Gehrke et al, even with multi-stage ultrasonic cleaning, which is also consistent with the current measurements [ 2 , 5 ].…”
Section: Discussionsupporting
confidence: 91%
“…The highest surface cleanliness was achieved by chemothermal ultrasonic cleaning (CP, FP and UD) and argon-oxygen plasma treatment (PL) (Table 4, Table 5). These results are consistent with Canullo et al and Gehrke et al, who also reported superior ultrasonic cleaning for residual and bacterial decontamination (2,5,37). The most signi cant differences in contamination levels were found between the FP/NC, CP/NC, UD/NC, and PL/NC groups.…”
Section: Discussionsupporting
confidence: 90%
“…While sterilisation does have an effect on the mechanical properties of the implant surface, titanium implants are particularly susceptible to this effect. There are studies that have shown that the sterilisation process in the autoclave is responsible for the alteration in the mechanical characteristics that leads to plastic deformation in steel materials when they are subjected to high levels of stress [ 26 , 27 ,28]. Mathew et al examined the mechanical properties of titanium and stainless steel mini-plates before and after repeated autoclaving cycles and found that the values were not constant.…”
Section: Discussionmentioning
confidence: 99%