2018
DOI: 10.1007/s10877-018-0231-3
|View full text |Cite|
|
Sign up to set email alerts
|

Mindray 3-directional NMT Module (a new generation “Tri-axial” neuromuscular monitor) versus the Relaxometer mechanomyograph and versus the TOF-Watch SX acceleromyograph

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
4
0

Year Published

2020
2020
2022
2022

Publication Types

Select...
2

Relationship

0
2

Authors

Journals

citations
Cited by 2 publications
(4 citation statements)
references
References 21 publications
0
4
0
Order By: Relevance
“…The i-TOF device consists of a triaxial accelerometer, two poles for nerve stimulation, and wireless connection on a smartphone by application. another stand-alone device, the TOF WATCH SX, whose efficacy has already been shown by several works [12][13][14][15] . In our study, we mainly found an overlap between the two devices in recorded mean values of TOF ratio, TOF count, and PTC analyzed at time intervals for every phase of general anesthesia, although the i-TOF tends to an underestimation compared to TOF WATCH SX.…”
Section: Discussionmentioning
confidence: 85%
See 1 more Smart Citation
“…The i-TOF device consists of a triaxial accelerometer, two poles for nerve stimulation, and wireless connection on a smartphone by application. another stand-alone device, the TOF WATCH SX, whose efficacy has already been shown by several works [12][13][14][15] . In our study, we mainly found an overlap between the two devices in recorded mean values of TOF ratio, TOF count, and PTC analyzed at time intervals for every phase of general anesthesia, although the i-TOF tends to an underestimation compared to TOF WATCH SX.…”
Section: Discussionmentioning
confidence: 85%
“…We compared i-TOF with the more studied and validated TOF WATCH SX (Organon, Ireland) [12][13][14][15] , using both simultaneously on 51 patients undergoing general anesthesia.…”
Section: Introductionmentioning
confidence: 99%
“…Although whether the onset time of the adductor pollicis muscle is slower than that of the diaphragm is still in dispute, the recovery time of the adductor pollicis muscle is indeed slower than the diaphragm when NMB was administered [ 15 , 65 ]. Therefore, this muscle gained enough attention from scholars when considering new techniques for neuromuscular monitoring [ 66 , 67 ].…”
Section: The Development Of Pmg-based Equipment For Perioperative Neu...mentioning
confidence: 99%
“…Despite its universal utilization in clinical settings, ACC is susceptible to artifacts and will probably generate an overrated block degree [ 12 , 14 ]. A conventional ACC-based apparatus records the monaxial movement of the target muscle; therefore, signal quality relies heavily on the posture of the recording sites [ 15 ]. Recently, ACC-based equipment capable of estimating the multiplanar movement of the thumb has been developed.…”
Section: Introductionmentioning
confidence: 99%