Abstract:The objective was to investigate the applicability and limitations of an approach for estimating particle size from powder dissolution measurement using as little as 50 microg of sample in 1 ml of buffer solutions. The powder dissolution profiles of five sparingly-soluble drugs (hydrochlorothiazide, phenazopyridine hydrochloride, 2-naphthoic acid, indomethacin, and dipyridamole) were evaluated with a novel biexponential spherical particle equation and also the Wang-Flanagan spherical particle non-sink equation… Show more
“…However, the miniaturized-volume dissolution apparatus, µDISS Profiler (Pion Inc.), uses 1.0 mL working volumes and makes it practical to determine solubility (Avdeef et al [69]; Avdeef and Tsinman [70]; Tsinman et al [71]; Avdeef et al [72]; Fagerberg et al [73]). Since the instrument uses in situ fiber optic UV (diode array) detection, the progression of dissolution may be followed in real time, making it possible to characterize the (transient) solubility of different polymorphs/solvates, over periods of days, if necessary.…”
Section: Appendix -Brief Summary Of Methods Suitable For Measuring Eqmentioning
“…However, the miniaturized-volume dissolution apparatus, µDISS Profiler (Pion Inc.), uses 1.0 mL working volumes and makes it practical to determine solubility (Avdeef et al [69]; Avdeef and Tsinman [70]; Tsinman et al [71]; Avdeef et al [72]; Fagerberg et al [73]). Since the instrument uses in situ fiber optic UV (diode array) detection, the progression of dissolution may be followed in real time, making it possible to characterize the (transient) solubility of different polymorphs/solvates, over periods of days, if necessary.…”
Section: Appendix -Brief Summary Of Methods Suitable For Measuring Eqmentioning
“…Concentrations were determined by considering area-underthe-curve (AUC) in second derivative spectra, evaluated over a range of wavelengths. Interference due to background turbidity in powder measurements is minimized by this spectral method (11)(12)(13)(14).…”
Section: Miniaturized Bath and In Situ Fiber Optic Dip-probe Uv Detecmentioning
confidence: 99%
“…This is because h app << h disk for very small particles (12,(15)(16)(17)(18). With rotating disks, the diffusion layer thickness, h disk , is time independent.…”
Section: Curve Fitting Concentration-time Profiles Of Dissolving Powdmentioning
confidence: 99%
“…Thus, API-sparing methods are critical to early adoption of dissolution testing (11)(12)(13). Additionally, these methods need to be media-sparing as well, especially when simulated intestinal fluid or human intestinal fluid is used.…”
Section: Introductionmentioning
confidence: 99%
“…The miniaturized-apparatus disk IDR values were compared to those determined by traditional equipment (taken from several studies) yielding a correlation coefficient r 2 = 0.99 (n = 31) (11). The API concentrations in the miniaturized apparatus were monitored in real time by an in situ dip-probe fiber optic UV method (11)(12)(13)(14).…”
The results demonstrate that using potentially 10,000-fold less drug material does not sacrifice the quality of the measurement, and lends support to an earlier study that the disk IDR measurement may possibly serve as a surrogate for the BCS solubility classification.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.