Abstract:Purpose
Signal voids caused by metallic needles pose visualization and monitoring challenges in many MRI applications. In this work, we explore a solution to this problem in the form of an active shim insert that fits inside a needle and corrects the field disturbance (ΔB0) caused by the needle outside of it.
Methods
The ΔB0 induced by a 4 mm outside‐diameter titanium needle at 3T is modeled and a two‐coil orthogonal shim set is designed and fabricated to shim the ΔB0. Signal recovery around the needle is asse… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.