Modification of the Dialysate Port of Plasma Separators for Safe Blood Purification in Japan

Abstract: A fatal mix‐up of a hemofilter with a plasma separator occurred in 2011. The close resemblance between the two blood purification columns commonly used in Japan posed a fundamental risk for such mix‐ups. Both the in‐hospital case investigation committee and the relevant academic societies have independently proposed the modifications of the dialysate port (D port) of the plasma separator to avoid this type of misuse. To make these devices foolproof, medical professionals, including physicians and clinical engi… Show more