“…In TRS‐398, the position of the reference point for the parallel‐plate chamber was at the reference depth (
and on the inner surface of the entrance window, at the center of the window, whereas for the cylindrical chamber, it was at the effective point of measurement, 0.5
deeper than
. This is different for measurements using modified calibration procedure by Muir
17 and Pawiro et al.,
18 where the position of the reference point for the cylindrical chamber was with the central axis at
The measurement of
in the absolute dosimetry of the electron beam is calculated as follows:
where
is the depth at which dose falls 50% of the maximum for 10 cm × 10 cm field size. The types of linear accelerators,
and
used in the measurements are listed in Table 4.…”