Monte Carlo calculation of Spencer-Attix and Bragg-Gray stopping-power ratios of tissue-to-air for ISO reference beta sources — an EGSnrc study

Abstract: On-axis Spencer-Attix and Bragg-Gray stopping power ratios of tissue-to-air are calculated for 25 radiation fields of ISO beta sources 147Pm, 85Kr, 90Sr/90Y and 106Ru/106Rh using the EGSnrc-based Monte Carlo code system. These values are calculated at 5 and 70 μm depths in a unit density ICRU 4-element tissue phantom, separately, for beta and photon fields using the published phase-space data at different distances with or without beam flattening filter. The study shows that the calculated values of stopping p… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

0
3
0

Year Published

2023
2023
2023
2023

Publication Types

Select...
3
2

Relationship

0
5

Authors

Journals

citations
Cited by 5 publications
(3 citation statements)
references
References 22 publications
0
3
0
Order By: Relevance
“…From the two new radiation fields, the spectral information is freely available [6]. From these data, corresponding Spencer-Attix stopping power ratios, s t,a (l) SA , depending on the chamber depth, l, were calculated using the method described above [17], the results will be published in [18]. Their ratio to the Bragg-Gray stopping power ratio, s t,a (0) BG , see table A9 in appendix A, can be interpreted as correction factor k SA to apply the Spencer-Attix theory:…”
Section: Use Of the Spencer-attix Theory Via Correction Factor K Samentioning
confidence: 99%
“…From the two new radiation fields, the spectral information is freely available [6]. From these data, corresponding Spencer-Attix stopping power ratios, s t,a (l) SA , depending on the chamber depth, l, were calculated using the method described above [17], the results will be published in [18]. Their ratio to the Bragg-Gray stopping power ratio, s t,a (0) BG , see table A9 in appendix A, can be interpreted as correction factor k SA to apply the Spencer-Attix theory:…”
Section: Use Of the Spencer-attix Theory Via Correction Factor K Samentioning
confidence: 99%
“…The parameters c 0 , c 1 and c 2 obtained from the regression are listed for several sources in table C5 in appendix C for BPS1. Figure 8 shows the data from the literature [23,24] and the corresponding polynomial regressions according equation (15). For a chamber depth of 1000 µm this results in values for k SA of 1.0036, 1.0045, 1.0057 and 1.0074 for 147 Pm at a distance of 20 cm, 85 Kr at a distance of 30 cm, 90 Sr/ 90 Y at a distance of 30 cm and 106 Ru/ 106 Rh also at a distance of 30 cm, all with a beam flattening filter, respectively.…”
Section: Use Of the Spencer-attix Theory Via Correction Factor K Samentioning
confidence: 99%
“…The Spencer-Attix cavity theory is considered to be more accurate than the Bragg-Gray one as it accounts for the variation in the response measured as a function of cavity dimension whereas the Bragg-Gray theory does not. Therefore, the Spencer-Attix stopping power ratios depending on the chamber depth have been calculated [43,44]. Corresponding correction factors were determined [33,34] and added in ISO 6980-2.…”
Section: Inclusion Of a Correction Factor For Primary Dosimetry To Us...mentioning
confidence: 99%