MP14: Community paramedic point of care blood analysis: validity and usability testing of two commercially available devices

Abstract: Introduction: Community Paramedics (CPs) require access to timely blood analysis in the field to guide treatment and transport decisions. Point of care testing (POCT), as opposed to traditional laboratory analysis, may offer a solution, but limited research exists on CP POCT. The objective of this study is to compare the validity of two POCT devices (Abbott i-STAT® and Alere epoc®) and their use by CPs in the community. Methods: In a CP programme responding to 6,000 annual patient care events, a split sample v… Show more