2024
DOI: 10.1186/s13014-024-02431-8
|View full text |Cite
|
Sign up to set email alerts
|

MR compatible detectors assessment for a 0.35 T MR-linac commissioning

Michel Chea,
Mathilde Croisé,
Christelle Huet
et al.

Abstract: Purpose To assess a large panel of MR compatible detectors on the full range of measurements required for a 0.35 T MR-linac commissioning by using a specific statistical method represented as a continuum of comparison with the Monte Carlo (MC) TPS calculations. This study also describes the commissioning tests and the secondary MC dose calculation validation. Material and methods Plans were created on the Viewray TPS to generate MC reference data. … Show more