BurnerVarían Techtron 5-cm high solids slot burner for nitrous oxide-acetylene. External Plano-convex Supracil lens 5-cm diameter with 12.5-cm focal length stopped down to 2.6 cm. Polka dot beam splitter Optics 1 inch by 1.5 inches, Instrumentation Laboratories. Slits Jarrell-Ash Model 82-092 0-400µ dual unilateral entrance and exit slit assembly with straight jaws.Spectrometer Jarrell-Ash Model 82-020 0.5-m scanning Ebert Mount monochromator with Model 82-016 camera attachment and an 1180 grooves-per-millimeter grating blazed at 400 nm, providing a reciprocal linear dispersion of 1.6 nm/mm. Detectors Ultraviolet sensitized silicon vidicon tube assembly Model 1205 F, Princeton Applied Research Corp. Silicon Intensified Target vidicon tube assembly Model 1205D, with option-01 scintillator screen, Princeton Applied Research Corp. RCA 1P28 Photomultiplier tube, Radio Corporation of America Power supply High voltage dc supply, Model 412A, John Fluke Mfg. Co., Inc. Amplifier Keithley Model 417 High speed picoammeter, Keithley Instruments, Inc. OMA Optical Multichannel Analyzer Model 1205A, Princeton Applied Research Corp. Data Readout Tektronix Oscilloscope, Model 604 Monitor. Speedomax X/L 680 Dual Channel Strip Chart Recorder, Leeds and Northrup Co.Flow meters Brooks Full-view Rotameters calibrated for nitrous oxide and acetylene, Brooks Instrument Co. emission lines at 184.9, 253.6, 312.5, 365.0, 404.6, and 546.1 nm were sequentially monitored by both detection systems. Each spectral line was first observed on the vidicon tube. The intensity was adjusted, using a lens diaphragm and neutral density filters, to yield a real-time reading of 100 counts for the peak channel on the OMA. The entrance