Abstract: We proposed a near‐infrared gas sensing system based on mode‐locked cavity‐enhanced absorption spectroscopy. A distributed feedback fiber laser (DFB‐FL) with a central wavelength of 1550 nm was employed as the light source, coupling with a Fabry–Perot (F‐P) cavity with an effective optical path length of 42.7 m served as the gas cell. By utilizing a homemade mode‐locked electric circuit and Pound–Drever–Hall (PDH) technique, the mode‐locking between the DFB‐FL and F‐P cavity was achieved with long‐term fluctua… Show more