Noise Exposure of Radiographer Caused by Magnetic Resonance Imaging(MRI)

Abstract: This study aimed to propose the necessity and system establishment of noise reduction facility via evaluating noise level exposed by the radiographer due to MRI scan. Noise measurements were carried out using at S general hospital in Daejeon using 1.5 Tesla MRI (7 exams) and 3.0 Tesla MRI (16 exams), while using SC-804 noise meter. The measurement distance was from the soundproof door of the MRI room to the radiographer which measured 100cm, and the measurement height, the height to the radiographer's ears whe… Show more