2019
DOI: 10.1016/j.ultrasmedbio.2018.08.021
|View full text |Cite
|
Sign up to set email alerts
|

Non-linear Wave Propagation and Safety Standards for Diagnostic Ultrasound Devices

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
3
0
2

Year Published

2020
2020
2024
2024

Publication Types

Select...
8
2

Relationship

0
10

Authors

Journals

citations
Cited by 12 publications
(5 citation statements)
references
References 30 publications
0
3
0
2
Order By: Relevance
“…The International Electrotechnical Commission (IEC) 60601-2-37 standard indicates a maximum temperature of 43°C or a temperature rise of 10°C when the probe is in contact with the skin [39,40]. IEC 62359 regulates acoustic output exposure parameters, setting limits such as in situ spatial peak intensity lower than 720 mW/cm², the spatial pulse peak average intensity lower than 190 mW/cm² and the mechanical index and thermal index lower than 1.9 and 6.0, respectively [41][42][43]. The standard also outlines the methods to measure the exposure parameters, where a hydrophone is used in distilled water since performing these measurements in situ is impractical.…”
Section: Safety Determinationmentioning
confidence: 99%
“…The International Electrotechnical Commission (IEC) 60601-2-37 standard indicates a maximum temperature of 43°C or a temperature rise of 10°C when the probe is in contact with the skin [39,40]. IEC 62359 regulates acoustic output exposure parameters, setting limits such as in situ spatial peak intensity lower than 720 mW/cm², the spatial pulse peak average intensity lower than 190 mW/cm² and the mechanical index and thermal index lower than 1.9 and 6.0, respectively [41][42][43]. The standard also outlines the methods to measure the exposure parameters, where a hydrophone is used in distilled water since performing these measurements in situ is impractical.…”
Section: Safety Determinationmentioning
confidence: 99%
“…Almost all the erythrocytes retained their biconcave disc shape after ultrasound stimulation, which is within the range of ultrasound frequency (o5 MHz) in clinical settings, which also verified the biosafety of sonication. 47,48 Besides, the movements at an ultrasound frequency of 1.46 MHz were recorded in the simultaneous presence of erythrocytes and RBC-MP-microbowls (Movie S4, ESI †). The RBC-MP-microbowls moved normally in the environment, while no obvious movements of erythrocytes were observed, demonstrating the selective ultrasound propulsion towards RBC-MP-microbowls.…”
Section: Ultrasound Responsibility and Propulsion Performance Of Rbc-...mentioning
confidence: 99%
“…Proses sonikasi diterapkan dalam farmasi, makanan, kosmetik, air, pelapis, tinta, cat, pengerjaan logam, perawatan kayu, pengerjaan logam(37), nanokomposit, pestisida, bahan bakar, produk kayu dan banyak industri lainnya. Sonikasi juga dapat digunakan untuk mempercepat pembubaran, dengan cara memecah interaksi antarmolekul serta menganalisis dinamika molekular dan kinetika reaksi pada pembelahan molekul (38).Ini sangat berguna ketika tidak memungkinkan untuk mengaduk sampel, seperti tabung NMRdan juga dapat digunakan sebagai penyedia energi (39) [19] , 2012-2016 oksidasi [20] 2017-2018 sonochemic al [21] , PCR [22] 2019> langsung (AS) [16] , Depolimerisasi [17] , Degradasi [18] , Tes triple negative praktis tidak termasuk PJI(64).…”
Section: Sonikasi Memiliki Banyak Efek Baik Kimia Maupun Fisikefek unclassified