Introduction: Longitudinal monitoring of vital signs provides a method for identifying changes to general health in an individual and particularly so in older adults. The nocturnal sleep period provides a convenient opportunity to assess vital signs. Contactless technologies that can be embedded into the bedroom environment are unintrusive and burdenless and have the potential to enable seamless monitoring of vital signs. To realise this potential, these technologies need to be evaluated against gold standard measures and in relevant populations. Methods: We evaluated the accuracy of heart rate and breathing rate measurements of three contactless technologies (two under-mattress trackers: Withings sleep analyser (WSA) and Emfit QS (Emfit) and a bedside radar: Somnofy) in a sleep laboratory environment and assessed their potential to capture vital signs (heart rate and breathing rate) in a real-world setting. Data were collected in 35 community dwelling older adults aged between 65 and 83 years (mean ± SD: 70.8 ± 4.9; 21 men) during a one-night clinical polysomnography (PSG) in a sleep laboratory, preceded by 7 to 14 days of data collection at-home. Several of the participants had health conditions including type-2 diabetes, hypertension, obesity, and arthritis and ≈ 49% (n = 17) had moderate to severe sleep apnea while ≈ 29% (n = 10) had periodic leg movement disorder. The under-mattress trackers provided estimates of both heart rate and breathing rate while the bedside radar provided only breathing rate. The accuracy of the heart rate and breathing rate estimated by the devices was compared to PSG electrocardiogram (ECG) derived heart rate (beats per minute, bpm) and respiratory inductance plethysmography thorax (RIP thorax) derived breathing rate (cycles per minute, cpm). We also evaluated breathing disturbance indices of snoring and the apnea-hypopnea index (AHI) available from the WSA. Results: All three contactless technologies provided acceptable accuracy in estimating heart rate [mean absolute error (MAE) < 2.2 bpm and mean absolute percentage error (MAPE) < 5%] and breathing rate (MAE ≤ 1.6 cpm and MAPE < 12%) at 1 minute resolution. All three contactless technologies were able to capture changes in heart rate and breathing rate across the sleep period. The WSA snoring and breathing disturbance estimates were also accurate compared to PSG estimates (R-squared: WSA Snore: 0.76, p < 0.001; WSA AHI: 0.59, p < 0.001). Conclusion: Contactless technologies offer an unintrusive alternative to conventional wearable technologies for reliable monitoring of heart rate, breathing rate, and sleep apnea in community dwelling older adults at scale. They enable assessment of night-to-night variation in these vital signs, which may allow the identification of acute changes in health, and longitudinal monitoring which may provide insight into health trajectories.