Novel Technique for Ultra‐Trace Determination of Lead in Seawater Samples Using Gold Coated Tungsten Coil Atom Trap Hydride Generation Atomic Absorption Spectrometry

Muhammet Atasoy

Abstract: A novel and simple trap technique combining a gold‐coated tungsten coil with hydride generation atomic absorption spectrometry (HGAAS) were developed for ultra‐trace lead determination. The technique was based on first transporting the plumbane produced by the HG system, together with H2, to a trap externally heated to 92 °C, where it was collected for in‐situ preconcentration and then volatilized by raising the trap temperature to 510 °C. For a 120 s trapping time, the limit of detection (LOD) was 2.9 ng L−1,… Show more