NRC measurement set-up and preparatory work for CCT-K7.2021 key comparison of triple-point-of-water cells

Abstract: This paper describes the NRC measurement set-up and the preparatory work for the CCT-K7.2021 key comparison of triple-point-of-water (TPW) cells. The preparatory work at NRC (charged with piloting the key comparison) included a comprehensive testing of the measurement set-up, the definition of the new NRC national standard for the TPW temperature, and the evaluation of the uncertainty budgets relevant for this key comparison. The new NRC TPW national standard is composed of an ensemble of 10 fused silica TPW c… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1

Citation Types

0
5
0

Year Published

2022
2022
2024
2024

Publication Types

Select...
3
1

Relationship

1
3

Authors

Journals

citations
Cited by 4 publications
(5 citation statements)
references
References 5 publications
0
5
0
Order By: Relevance
“…Measurement setup. The new results presented in this paper were obtained using the measurement setup for the CCT-K7.2021 TPW key comparison which is described in detail in [8]. Briefly, for TPW resistance ratio measurements, we used a 25 Ω SPRT (Leeds and Northrup, s/n 1504288) and a 25 Ω standard resistor (Tinsley 5685A, s/n 274880) connected to a DC thermometry bridge (Measurements International MI6020T, s/n 1104110).…”
Section: Methodsmentioning
confidence: 99%
See 4 more Smart Citations
“…Measurement setup. The new results presented in this paper were obtained using the measurement setup for the CCT-K7.2021 TPW key comparison which is described in detail in [8]. Briefly, for TPW resistance ratio measurements, we used a 25 Ω SPRT (Leeds and Northrup, s/n 1504288) and a 25 Ω standard resistor (Tinsley 5685A, s/n 274880) connected to a DC thermometry bridge (Measurements International MI6020T, s/n 1104110).…”
Section: Methodsmentioning
confidence: 99%
“…There was no uncertainty budget compiled for the measurements completed in 2007 and 2013, so one can use the NRC CCT-K7 (2002-2005) uncertainty (k = 2) of 29 μK (see table 18 in [9]) as a guiding value. The uncertainty (k = 2) in measuring the temperature difference between a TPW cell and NRC national reference in 2018 and 2021 was 46 μK [16] and 39 μK [8], respectively. Measurement method.…”