To compare quantitative changes in macular parameters in diabetic patients detected by two optical coherence tomography angiography (OCTA) instruments. Methods 80 phakic eyes were classified as no diabetes, diabetes without diabetic retinopathy (DR), mild non-proliferative diabetic retinopathy (NPDR), and severe NPDR or proliferative DR (PDR). OCTA was performed using devices from two manufacturers (Zeiss and Heidelberg). Superficial and deeper vascular skeleton density (SVSD, DVSD), superficial and deeper vessel area density (SVAD, DVAD), choriocapillaris flow voids (CCFV), and choroidal flow voids (CFV) were calculated. Inter-device comparisons were performed using the size comparison index (SCI) and the discrepancy index (DI). Results The two devices were inconsistent in SVSD, DVSD, DVAD, CCFV and CFV parameters (all P < 0.05). In addition, the SCI was positive for DVAD (all P < 0.001) and negative for SVSD, DVSD, CCFV and CFV in all groups (all P <0.001), except for DVSD in severe NPDR or PDR. The discrepancy index was not significantly different among groups for SVD, SPD, DVD, DPD and CFV (all P> 0.05). The mean DI of CCFV was statistically different between the four groups (P < 0.001).