Background
The Viabahn endoprosthesis has become a vital option for endovascular therapy, yet there is limited long-term data on its effectiveness for peripheral aneurysm repair. This study aimed to evaluate the safety, technical and clinical success, and long-term patency of the Viabahn endoprosthesis for treating femoropopliteal aneurysms.
Methods
This retrospective tertiary single-center study analyzed patients who underwent a Viabahn endoprosthesis procedure for femoropopliteal aneurysm repair from 2010 to 2020. Intraoperative complications, technical and clinical success rates, and major adverse events (MAE, including acute thrombotic occlusion, major amputation, myocardial infarction, and device- or procedure-related death) at 30 days were assessed. Incidence of clinically-driven target lesion revascularisation (cdTLR) was noted. Patency rates were evaluated by Kaplan–Meier analysis.
Results
Among 19 patients (mean age, 72 ± 12 years; 18 male, 1 female) who underwent aneurysm repair using the Viabahn endoprosthesis, there were no intraoperative adverse events, with 100% technical and clinical success rates. At the 30-day mark, all patients (19/19, 100%) were free of MAE. The median follow-up duration was 1,009 days [IQR, 462–1,466]. Popliteal stent graft occlusion occurred in 2/19 patients (10.5%) after 27 and 45 months, respectively. Consequently, the primary patency rates were 100%, 90%, 74% at 12, 24, and 36–72 months, respectively. Endovascular cdTLR was successful in both cases, resulting in sustained secondary patency at 100%.
Conclusion
The use of Viabahn endoprostheses for femoropopliteal aneurysm repair demonstrated technical and clinical success rates of 100%, a 0% 30-day MAE rate, and excellent long-term patency.